Home > Products > Melting furnace
35KW (30-80kHz) Induction Heating System with Timer Control

35KW (30-80kHz) Induction Heating System with Timer Control

    CY-SP-30 induction heater features the latest inverting technology so as to make itself light and convenient. Its 30KW Induction Heater is designed for the material research lab to quickly melt and perform heat treating to metallic materials.

CY-SP-30 induction heater features the latest inverting technology so as to make itself light and convenient. Its 30KW Induction Heater is designed for the material research lab to quickly melt and perform heat treating to metallic materials.

Technical parameters of induction heating system:

Induction Heater

  • Working Voltage: 220V AC, 50/ 60Hz, Three Phases ( 380VAC, 3phase available upon request )

  • Working Current: 15~110 Amps (125A breaker required)

  • Max. Input Power: 35 KW 

  • Output Frequency: 30 - 80 kHz 

  • Duty Cycle: 80%

  • Induction Coil: 150mm(OD) x 140mm(ID) x 65mm(H) 

  • Air Cooling: Rear Fan

  • Protection: Automatic water pressure, over temperature, and over power protection

  • Refractory Assembling: Complete refractory assembling made of alumina, including bottom crucible support, crucible holder, and cover

Timer Control

Automatic timer control or Manual control selectable

  • Manual Control: by Foot Pedal

  • Automatic Heating Time: 1-99 Seconds

  • Automatic Dwell Time: 1-99 Seconds

  • Automatic Cooling Time: 1-99 Seconds

Temp. Controller & Thermocouple (optional)

  • Precision digital temperature controller with Automatic PID tuning

  • 30 segments programmable for controlling heating and cooling rate as well as dwell time.

  • Overheating & thermocouple failure protection.

  • +/- 3℃ temperature accuracy. 

  • +/- 3℃ temperature accuracy. 

  • Thermocouple: C type  (1/4"x10"L)

  • Continuous Working Temp. Range: 1000℃~1900℃

  • Max. Working Temperature: 2000℃ (for less than an hour)

  • Max. Heating Rate: 

    •  4℃/second (1000℃-1200℃)

    • 3℃/second (1200℃~1500)

    •  1℃~2℃/second (1500℃~2000℃)

Water Cooling

  • Flowrate: >10 L/Min

  • Pressure: ≥0.2 MPa (29 Psi)

  • Tap water is acceptable, but using Recirculating Water Chiller is highly recommended. Protection: Overpressure and over temperature switch

  • Water Chiller (optional)

    • Temperature range: 5 - 30℃

    • Water Flow Rate: > 58 L/m

    • Note: It is recommended to use a specific coolant to get the best performance.

Warranty

  • One year limited warranty with lifetime support

  • Consumable parts such as processing tube, thermocouple, refractory liners, o-rings, and heating coil are not covered by the warranty


Disclaimer: The product introduction content (including product images, product descriptions, technical parameters, etc.) of this site is for reference only. As the update may not be timely, there will be some differences between the content and the actual situation. Please contact our sales staff to confirm. The information provided on this site does not constitute any offer or promise, and the company will periodically improve and modify the website information, without prior notice, please understand.

Contact Us
  • E-mail: cysi@cysi.wang
  • Tel: +86 371 5519 9322
  • Fax: +86 371 8603 6875
  • Add: No. 820, 8th Floor, 1st Unit, 9th Block, Cuizhu Street, High-Tech Zone, Zhengzhou, Henan, China




Follow Us

Copyright © Zhengzhou CY Scientific Instrument Co., Ltd. All Rights Reserved   

| Sitemap |       Technical Support: